Sartolab® P20, 0.2 µm, PES, Luer Lock
The Sartolab® P20 is a pressure filtration unit for sterile filtration or clarification of media, aqueous solutions and solvent mixtures in batches ranging from 100 ml to 10 L. For media that contain sera and difficult-to-filter solutions, a Sartolab® P20 Plus unit with an incorporated prefilter is available.
- Price (CNY):
- CNY949.75
- Item No.:
- 18090----------D
- Availability:
- pcs
*Custom/bulk order quotes are provided within 24 hours of request.
Choose your Sartolab® P20 Pressure Filters
Membrane: 0.2 µm polyethersulfone
Housing: Transparent polycarbonate
Housing diameter: 61 mm
Inlet: Female Luer-Lock
Outlet: Male Luer-Lock
Filtration Area: 21 cm²
Holdup Volume: 1 mL
Recommended Max.Inlet Pressure: 4 bar
Housing Burst Pressure: > 5 bar
pH Range: 1-10 Box à 10 pieces, individually packed and sterile
Applications
-
ApplicationClarification
-
Application AreaSterile filtration
Components
-
Filling BellNo
-
PrefilterNo
Connections (Physical)
-
Connector InletFemale Luer Lock
-
Connector OutletMale Luer Lock
Dimensions
-
Filter Diameter61 mm
-
Filtration Area20 cm²
-
Housing Dimensions61 x 31 mm
Features
-
Venting FilterNo
General Specifications
-
Flow Rate≥ 400 mL/min at 1 bar
-
Total ThroughputUp to 5 L
-
VentNo
-
pH Range1.0 – 10.0
Materials of Construction
-
Housing MaterialPolycarbonate (PC)
-
Material PrefiltersN/A
-
Membrane MaterialPolyethersulfone (PES)
Operating Conditions
-
Maximum Allowable Working Temperature50 °C (122 °F)
-
Operating Pressure4 bar
Physicochemical Information
-
Bubble Point≥ 3.2 bar (46 psi) with water
-
Burst Pressure≥ 0.5 bar (72.5 psi)
-
Non-Fiber ReleasingThis filter product complies with the title 21 of the Code of Federal Regulations (CFR), section 210.3(b)(6) and 211.72.
-
Pore Size0.2 µm
-
WettabilityHydrophilic
Product Information
-
BrandSartolab® P20
-
Device TypePressure filters without prefilter
-
PackagingSingle pouch packed
-
Sample Type
-
Aqueous
- Solvents according to chemical compatibility
-
Aqueous
Sterility/Purity
-
Delivery ConditionSterile
-
Sterilization MethodEthylene oxide
Volume
-
Hold-up Volume1 mL before bubble point
-
VolumeUp to 5 L